It is highly recommended that you also purchase the SPM-CS 3L calibration syringe, in order to ensure accurate calibration of your spirometer.
The SpiroCare unit is an accurate and hygienic spirometer. It features 7” color touch screen, provides FVC/SVC/MVV/post-bronchodilator tests, candle blowing-out animation to encourage patient success, accurate measurement from disposable mouthpiece with patented-sensor method, prints PFT test report with interpretation, and a 4-year warranty.
For data management, it saves up to 200 tests internally and exports saved data to USB flash drive, in JPEG/PDF/DCM file formats.
Performance
- FVC (Forced Vital Capacity), SVC (Slow Vital Capacity), and MVV (Maximum Voluntary Maximum) test.
- Pre- and post-bronchodilator comparison.
- Accurate measurement by using disposable mouthpiece with built-in sensor.
- PFT test report with interpretation.
User-Friendliness
- 7” color screen for displaying measurement in real time.
- Touch interface for easily entering patient information.
- Built-in printer and rechargeable battery.
- USB barcode reader and keyboard can be attached through USB ports.
- Disposable mouthpiece with patented sensing method is hygienic and eliminates the hassle of cleaning.
Data Management
- Saves up to 200 tests internally.
- Exports the saved data to USB flash drive as JPEG/PDF/DCM file format.
Special Advantage
- 3-year warranty
Includes:
SpiroCare Main Unit with Spirometer Handle, 1 Medi-Graph Paper Roll, 20 Disposable Mouthpieces, 1 Nose Clip, 1 Mouthpiece Adapter, Hospital Grade Power Cord, Operation Manual, Diagnostics Guide Book and 3-Year Warranty Card. WiFi USB Dongle and USB Flash Drive with BMS-Plus Software Included.
Option:
Cart
Warranty:
3-Year Warranty
Calibration
It is recommended to calibrate a spirometer with Bionet 3L calibration syringe periodically.
Our calibration recommendations are based upon guidelines from the American Thoracic Society (ATS) and the European Respiratory Society (ERS)
- The volume accuracy of the spirometer must be checked at least daily with a single discharge of a 3-L calibrated syringe.
- At least quarterly, volume spirometers must have their calibration checked over their entire volume range using a calibrated syringe or an equivalent volume standard. The measured volume should be within 3.5% of the reading or 65 mL, whichever is greater. This limit includes the 0.5% accuracy limit for a 3-L syringe.
Specification:
| FEATURE | SPECIFICATION |
|---|---|
| Dimension | 296 (W) x 305.5 (H) x 34 (D) mm, approx. 250 kg |
| Measuring Values | – FVC: FVC, FEV 1.0, FEV 1/FVC, FEF 0.2-1.2l, fef 25-75%, FEF 75-85%, PEF, FEF 25%, FEF 50%, FEF 75%, FIVC, FEV6, PEFT, FET100%, Extrapolated Volume – SVC: SVC, ERV, IRV, TV, EC – MVV: MVV, FB, TV |
| Presentation | Flow volume loop MVV: MVV, FB, TV Measurement values table |
| Measuring Range | Flow: 0 to ± 14 L/s Volume: 0 to ± 12 L |
| Measuring Method | Differential pressure method |
| Prediction equation | Morris-Polgar, ECCS-Quanjer, Knudson-ITS, Pereira, Korea CJK |
| Sample rate | 200 samples/sec |
| Flow impedance | < 0.2 mbar s/l at 12 l/s |
| Measuring accuracy | Complies with ATS (American Thoracic Society) |
| Environment | Ambient temperature: 15 to 40°C (59 to 104°F) Relative humidity: 10 to 90% (non-condensing) Atmospheric pressure: 700 to 1060 hPa |
| Standard accessory | Disposable mouthpiece 20 ea. Nose clip 1 ea. Adapter 1 ea. Handle dock 1 ea. Operation manual 1 ea. Diagnosis guide 1 ea. Power cord 1 ea. |
* Specifications can be changed without prior notification.